Fuji Labs ZSP instruction manual Cal

Page 21

[When adjusting for each range]

(Simultaneous calibration of H and L ranges has been factory-set).

Perform setting in the following way.

NO

 

By pressing

CAL

key while the highest digit of the

 

main indication for the 1st component is flickering,

 

then ‘

’ will appear on the main indication and

 

 

‘ ’ will appear on the sub indication.

NO

 

 

 

 

CAL

Press key and will change to .

= Meaning of set values =

: The calibration is valid only for the selected range, and zero and span calibration can

be made independently for each range.

: By conducting zero and span calibration

for one range, the calibration will be done

NO

NO

ENT

automatically for the other ranges as well.

When the selection has been finished, press the ENT key. Setting operation is completed.

– 13 –

Image 21
Contents Infrared GAS Analyzer Request IntroductionProhibition Iii Prohibition Others Contents Vii Outline SpecificationsComposition PreparedInstallation Selection of installation siteSelection of gas sampling location Connection of cables Installation workPiping procedure Supply of external power Injection of clean waterDescription of Each Part Preparation for OperationSetting of temperature inside locker Airtightness testWarm-up time SO2 INDICATION/OPERATION PanelThis panel consists of the following functions Function and indication of indication/operation panelGeneral operation of indication/operation panel Advice on OperationCalibration Kinds of standard gasPreparation of standard gas Pressure regulation for standard gas cylinder Func CAL SET ENT Calibration concentration setting except zirconia O2 sensorCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operation Operating procedureRoutine maintenance and inspection MaintenanceConditioner Other maintenance and inspection Periodic maintenance and inspectionCentration is measured in the setting range of 0 to 200ppm or more at 0.5L/minGas extractor type ZBAK1 Maintenance of sampling deviceMembrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Others, major component figures Maintenance of zirconia O2 sensorStandard accessories Standard Accessories and SparesCode for preparation of 1-year spares and supplies Spares for 1 year optionalOutline Processing of Heating Pipe EndsEnd processing method Grid 300mm 40mm Grid area 150 mm Clamp support Check after pipingSpecifications of heating pipe Specifications of tube traceTroubleshooting of instrument TroubleshootingError codes and remedies Power supply Ambient temperaturePower consumption Mass weightStandard requirements For sample gas Month/NO2, 100ppmNumber of output points and kinds of outputs Moving average output Block diagramOrdering information Automatic calibrationCalibration tolerance Automatic calibration output hold Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For indoor installation type For outdoor installation typeOnly specified outputs are wired Connection DiagramHead Office Instrumentation Div International Sales Dept