Fuji Labs ZSP instruction manual Prohibition

Page 5

Caution on use

Prohibition

Do not insert a rod or your finger into the fan (electronic cooler at ceiling), otherwise injury may be caused by the rotating blades.

Do not touch metal, finger, etc. to the input and output terminal block of analyzer. Otherwise, shock hazard or injury may be caused.

Smoking and use of fire/flame is prohibited in the vicinity of the gas analyzer. Otherwise, a fire may occur.

A water-filled container such as a bucket should not be placed on the roof of the gas analyzer having indoor specifications. If such a container turns over, spilled water may cause shock hazard or burning inside the analyzer.

Caution on maintenance / inspection

Danger

When handling standard gas such as calibration gas, read the instruction manual for the standard gas for correct handling. Special care should be taken when handling hazardous gas such as CO gas and the like to prevent poisoning.

During maintenance or inspection, be sure to turn ON the ventilation fan in the analyzer to prevent poisoning due to gas leakage.

When the gas filter of the analyzer is to be replaced or the washing unit needs maintenance, be sure to close the calibration gas valve. Especially, if the calibration gas valve is located at the sample gas inlet, the replacing or washing shall be done after closing the calibration gas valve to prevent poison- ing or other trouble.

Caution

The gas converter is heated to about 220°C. When the catalytic agent needs to be replaced, turn OFF the power for the converter and use heat-proof gloves about 30 minutes later.

When the fuse is blown off, check the cause of it and replace it with the same type of fuse having the same capacity to prevent electric shocks or other trouble.

When the analyzer is provided with a power leak relay, press the test button to confirm proper opera- tion before making maintenance or inspection to prevent electric shocks.

Be sure to lock the door. Do not allow unauthorized persons to use the door key. Do not touch the door carelessly to prevent electric shocks and other trouble.

The following items must be observed to prevent electric shocks or injury and to ensure safety.

Take off metallic objects such as wrist watch before starting the work.

Do not touch the device with wet hands.

– iv –

Image 5
Contents Infrared GAS Analyzer Request IntroductionProhibition Iii Prohibition Others Contents Vii Outline SpecificationsComposition PreparedSelection of gas sampling location InstallationSelection of installation site Connection of cables Installation workPiping procedure Supply of external power Injection of clean waterDescription of Each Part Preparation for OperationWarm-up time Setting of temperature inside lockerAirtightness test SO2 INDICATION/OPERATION PanelThis panel consists of the following functions Function and indication of indication/operation panelGeneral operation of indication/operation panel Advice on OperationPreparation of standard gas CalibrationKinds of standard gas Pressure regulation for standard gas cylinder Func CAL SET ENT Calibration concentration setting except zirconia O2 sensorCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operation Operating procedureRoutine maintenance and inspection MaintenanceConditioner Other maintenance and inspection Periodic maintenance and inspectionCentration is measured in the setting range of 0 to 200ppm or more at 0.5L/minGas extractor type ZBAK1 Maintenance of sampling deviceMembrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Others, major component figures Maintenance of zirconia O2 sensorStandard accessories Standard Accessories and SparesCode for preparation of 1-year spares and supplies Spares for 1 year optionalOutline Processing of Heating Pipe EndsEnd processing method Grid 300mm 40mm Grid area 150 mm Clamp support Check after pipingSpecifications of heating pipe Specifications of tube traceTroubleshooting of instrument TroubleshootingError codes and remedies Power supply Ambient temperaturePower consumption Mass weightNumber of output points and kinds of outputs Standard requirements For sample gasMonth/NO2, 100ppm Moving average output Block diagramOrdering information Automatic calibrationCalibration tolerance Automatic calibration output hold Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For indoor installation type For outdoor installation typeOnly specified outputs are wired Connection DiagramHead Office Instrumentation Div International Sales Dept