8.7.3Auto calibration cycle setting

The calibration cycle is settable in either ‘days’s or ‘hours’.

Press the FUNC key at the calibration start time indica-

tion and ‘ ’ will appear.

The AUTO CAL LED of the function indicating lamps will flicker.

Press the key and the auto calibration cycle is settable. The highest digit on the sub indication will flicker. Select

either ‘days’ or ‘hours’ by pressing the key.

’ appears on the sub indication when selecting ‘days’ and ‘’ appears when selecting ‘hours’.

Press the key and the calibration cycle and be set in the lower digits of the sub indication.

The numeric value is incremented by pressing the key.

Press the key and the digit can be selected.

Day is settable in a range of 1 to 7 days. Hour is settable in a range of 1 to 99 hours.

Note

If ‘0 0’ (hours) is set, it will be automatically set to ‘1’(hours).

Press the ENT key when the calibration cycle has been

set. The data are saved.

NO

FUNC

NO

AUTO CAL

( )

NO

( )

NO

ENT

NO

– 22 –

Image 30
Contents Infrared GAS Analyzer Introduction RequestProhibition Iii Prohibition Others Contents Vii Composition SpecificationsOutline PreparedInstallation Selection of installation siteSelection of gas sampling location Installation work Connection of cablesPiping procedure Description of Each Part Injection of clean waterSupply of external power Preparation for OperationSetting of temperature inside locker Airtightness testWarm-up time INDICATION/OPERATION Panel SO2Function and indication of indication/operation panel This panel consists of the following functionsAdvice on Operation General operation of indication/operation panelCalibration Kinds of standard gasPreparation of standard gas Pressure regulation for standard gas cylinder Calibration concentration setting except zirconia O2 sensor Func CAL SET ENTCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operating procedure OperationMaintenance Routine maintenance and inspectionConditioner Centration is measured in the setting range of 0 to Periodic maintenance and inspectionOther maintenance and inspection 200ppm or more at 0.5L/minMaintenance of sampling device Gas extractor type ZBAK1Membrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Maintenance of zirconia O2 sensor Others, major component figuresStandard Accessories and Spares Standard accessoriesSpares for 1 year optional Code for preparation of 1-year spares and suppliesProcessing of Heating Pipe Ends OutlineEnd processing method Grid 300mm 40mm Grid area 150 mm Check after piping Clamp supportSpecifications of tube trace Specifications of heating pipeTroubleshooting Troubleshooting of instrumentError codes and remedies Power consumption Ambient temperaturePower supply Mass weightStandard requirements For sample gas Month/NO2, 100ppmNumber of output points and kinds of outputs Block diagram Moving average outputCalibration tolerance Automatic calibration output hold Automatic calibrationOrdering information Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For outdoor installation type For indoor installation typeConnection Diagram Only specified outputs are wiredHead Office Instrumentation Div International Sales Dept