Fuji Labs ZSP instruction manual Single-component analyzer

Page 25

8.6Span calibration (low concentration point calibration in case of zirconia O2 sensor)

Carry out span calibration by flowing calibration gas at the concentration set as a span value.

Flow the calibration gas (0.3 N/min.) by adjusting the needle valve of pressure regulator. Wait until the indica- tion stabilizes, and then conduct span calibration.

Press the SPAN key and the SPAN key lamp flickers.

In the case of a multi-component analyzer, the main indication will flicker for the components which can be calibrated.

Press the COMP key and the settable gas compo- nent will change.

When the gas component has been determined,

press the keys and select a range.

Note: The lamp will not flicker in the case of a

single-component analyzer.

Next press the CAL key and conduct span calibration. The CAL key lamp lights steadily during calibration.

Since the measuring status is resumed after calibration, continue the calibration for each component and each range.

NO

 

 

 

 

SPAN

 

 

NO

 

 

 

 

 

 

Flickers

 

 

ENT

ZERO SPAN CAL

(

COMP

)

 

NO

 

 

 

 

 

 

Flickers

 

 

ENT

ZERO SPAN CAL

(In case of a multi component analyzer)

 

CAL

 

 

NO

 

 

 

 

 

 

lights

ENT ZERO SPAN CAL

ENT

ZERO SPAN CAL

( Lamp is turned off after calibration )

Note: On the low range of zirconia type O2 sensor, when zero gas is calibrated with 9 to 10 Vol. % O2, the zero and span calibrations interfere with each other. So, each calibration should be re- peated a few times.

When the function for automatic calibration is provided, the solenoid valve turns on and the span gas for the selected component flows upon pressing the SPAN key.

– 17 –

Image 25
Contents Infrared GAS Analyzer Request IntroductionProhibition Iii Prohibition Others Contents Vii Outline SpecificationsComposition PreparedSelection of installation site InstallationSelection of gas sampling location Connection of cables Installation workPiping procedure Supply of external power Injection of clean waterDescription of Each Part Preparation for OperationAirtightness test Setting of temperature inside lockerWarm-up time SO2 INDICATION/OPERATION PanelThis panel consists of the following functions Function and indication of indication/operation panelGeneral operation of indication/operation panel Advice on OperationKinds of standard gas CalibrationPreparation of standard gas Pressure regulation for standard gas cylinder Func CAL SET ENT Calibration concentration setting except zirconia O2 sensorCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operation Operating procedureRoutine maintenance and inspection MaintenanceConditioner Other maintenance and inspection Periodic maintenance and inspectionCentration is measured in the setting range of 0 to 200ppm or more at 0.5L/minGas extractor type ZBAK1 Maintenance of sampling deviceMembrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Others, major component figures Maintenance of zirconia O2 sensorStandard accessories Standard Accessories and SparesCode for preparation of 1-year spares and supplies Spares for 1 year optionalOutline Processing of Heating Pipe EndsEnd processing method Grid 300mm 40mm Grid area 150 mm Clamp support Check after pipingSpecifications of heating pipe Specifications of tube traceTroubleshooting of instrument TroubleshootingError codes and remedies Power supply Ambient temperaturePower consumption Mass weightMonth/NO2, 100ppm Standard requirements For sample gasNumber of output points and kinds of outputs Moving average output Block diagramOrdering information Automatic calibrationCalibration tolerance Automatic calibration output hold Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For indoor installation type For outdoor installation typeOnly specified outputs are wired Connection DiagramHead Office Instrumentation Div International Sales Dept