8.5Zero calibration (air point calibration in case of zirconia O2 sensor)

This is used for adjusting the zero point.

Flow the zero gas (0.3N/min.) by adjusting the needle valve of pressure regulator and wait until the indica- tion stabilizes.

When the Zirconia O2 sensor is used, use air.Note)

When the indication has stabilized, select the measur-

ing range to be calibrated by using the keys.

When using a multi-component analyzer, all the components undergo zero calibration simulta- neously at the selected range.

Press the ZERO key, and ZERO key lamp will flicker.

Next press the CAL key and conduct zero calibra- tion.

The CAL key lamp lights steadily during calibra- tion.

NO

ZERO

NO

Flickers

ENT ZERO SPAN CAL

CAL

NO

Lights

ENT ZERO SPAN CAL

ENT ZERO SPAN CAL

( Lamp is turned off after calibration )

Since the measuring status is resumed after the cali- bration, carry out the calibration for other ranges if necessary.

Note: When using different gases for the low and high ranges, apply 9 to 10 Vol. % O2/N2 which has been set for the low range.

When the function for automatic calibration is provided, the solenoid valve turns on and cali- bration gas flows upon pressing the ZERO key.

– 16 –

Image 24
Contents Infrared GAS Analyzer Introduction RequestProhibition Iii Prohibition Others Contents Vii Specifications OutlineComposition PreparedInstallation Selection of installation siteSelection of gas sampling location Installation work Connection of cablesPiping procedure Injection of clean water Supply of external powerDescription of Each Part Preparation for OperationSetting of temperature inside locker Airtightness testWarm-up time INDICATION/OPERATION Panel SO2Function and indication of indication/operation panel This panel consists of the following functionsAdvice on Operation General operation of indication/operation panelCalibration Kinds of standard gasPreparation of standard gas Pressure regulation for standard gas cylinder Calibration concentration setting except zirconia O2 sensor Func CAL SET ENTCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operating procedure OperationMaintenance Routine maintenance and inspectionConditioner Periodic maintenance and inspection Other maintenance and inspectionCentration is measured in the setting range of 0 to 200ppm or more at 0.5L/minMaintenance of sampling device Gas extractor type ZBAK1Membrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Maintenance of zirconia O2 sensor Others, major component figuresStandard Accessories and Spares Standard accessoriesSpares for 1 year optional Code for preparation of 1-year spares and suppliesProcessing of Heating Pipe Ends OutlineEnd processing method Grid 300mm 40mm Grid area 150 mm Check after piping Clamp supportSpecifications of tube trace Specifications of heating pipeTroubleshooting Troubleshooting of instrumentError codes and remedies Ambient temperature Power supplyPower consumption Mass weightStandard requirements For sample gas Month/NO2, 100ppmNumber of output points and kinds of outputs Block diagram Moving average outputAutomatic calibration Ordering informationCalibration tolerance Automatic calibration output hold Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For outdoor installation type For indoor installation typeConnection Diagram Only specified outputs are wiredHead Office Instrumentation Div International Sales Dept