8.4Calibration concentration setting (zirconia O2 sensor)

A zirconia O2 sensor sets the zero (air) point (span point stipulated in the laws for measurement and weighing) of the O2 analyzer. For the density of calibration gas to be used, refer to “8.2 Preparation of standard gas”.

Setting of calibration concentration for zero (air) point (zero stipulated in the laws for measurement and weighing)

By pressing the FUNC key in the measuring status,

the calibration concentration previously set for the zero (air) point will appear on the main indication.

The CAL SET LED of the function indicating lamps will flicker.

By pressing the key, the higher digit of the main indication of the O2 analyzer will flicker and zero point setting is enabled.

Press the keys to select a range.

The numeric value will be incremented by pressing

key. The digit to be set is selected by pressing the

key.

Setting ranges are 20.2 to 21.9 Vol. % for high range and 7.0 to 21.9 Vol. % for law range.

Set the air concentration for both high and low range.Note)

Press the ENT key after setting the zero point. Set- ting operation is completed.

O2

FUNC

O2

CAL SET

O2

O2

ENT

O2

Note: When using air for the high range and 9 to 10 Vol. % O2/N2 for the low range, set the density for each range.

When viewing from the zirconia O2 sensor side, the air point corresponds to the zero point. Therefore, calibration should begin with the air point.

Also, since the air point and low concentration point can be calibrated by the zero calibra- tion key and span calibration key, respectively, the air point and low concentration point are regarded as the zero and span in this manual.

– 14 –

Image 22
Contents Infrared GAS Analyzer Introduction RequestProhibition Iii Prohibition Others Contents Vii Composition SpecificationsOutline PreparedSelection of installation site InstallationSelection of gas sampling location Installation work Connection of cablesPiping procedure Description of Each Part Injection of clean waterSupply of external power Preparation for OperationAirtightness test Setting of temperature inside lockerWarm-up time INDICATION/OPERATION Panel SO2Function and indication of indication/operation panel This panel consists of the following functionsAdvice on Operation General operation of indication/operation panelKinds of standard gas CalibrationPreparation of standard gas Pressure regulation for standard gas cylinder Calibration concentration setting except zirconia O2 sensor Func CAL SET ENTCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operating procedure OperationMaintenance Routine maintenance and inspectionConditioner Centration is measured in the setting range of 0 to Periodic maintenance and inspectionOther maintenance and inspection 200ppm or more at 0.5L/minMaintenance of sampling device Gas extractor type ZBAK1Membrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Maintenance of zirconia O2 sensor Others, major component figuresStandard Accessories and Spares Standard accessoriesSpares for 1 year optional Code for preparation of 1-year spares and suppliesProcessing of Heating Pipe Ends OutlineEnd processing method Grid 300mm 40mm Grid area 150 mm Check after piping Clamp supportSpecifications of tube trace Specifications of heating pipeTroubleshooting Troubleshooting of instrumentError codes and remedies Power consumption Ambient temperaturePower supply Mass weightMonth/NO2, 100ppm Standard requirements For sample gasNumber of output points and kinds of outputs Block diagram Moving average outputCalibration tolerance Automatic calibration output hold Automatic calibrationOrdering information Calibration cycle Once/day variable, up to 7 daysCode Symbols Diaglam Functions of individual components For outdoor installation type For indoor installation typeConnection Diagram Only specified outputs are wiredHead Office Instrumentation Div International Sales Dept