Hex. socket head cap screw
Cap A
Seat valve

(4) Replacement of the valve and the diaphragm of diaphragm type aspirator

Replace the seat valve (turn 90° in six months and replace annually) and the diaphragm (replace annually) according to the following procedure.

(a)Turn off the power supply of the aspirator and remove the inlet and outlet piping connected to the aspirator.

(b)Remove the four hexagon socket head cap screws. Then cap A and cap B are separated and the valve can be removed.

(c) Turn the diaphragm counterclockwise with your hand and remove it.

(d) Place a new diaphragm to the position and mount it by turning it clockwise until it stops.

Note) When mounting, check that the dia- phragm is securely engaged in the screw section of the arm slot. Other- wise malfunction may result.

(e)Place the valve on cap B and turn it 90. Check that the notches of cap A and cap B are aligned, and then tighten the four hexagon socket head cap screws.

Note) Since complex valve seat is adopted, re- move the pin on the underside of the valve from the hole to which the pin has been inserted and reinsert it to the other hole.

(f)Turn on the power of the aspirator in the inter- face module. Check that it does not generate abnormal sound. Check also with your hand that the valve is normally operated, that is, air is as- pirated from the IN side and discharged from the OUT side. If any abnormality is observed, repeat the procedure from the beginning.

(g)After checking that the aspirator is normally op- erated, turn off the power of the aspirator in the interface module, and return the piping to the original position.

Note) Do not apply force abruptly to Rc1/8 screw part when returning the piping to the original position.

Rc1/8

Cap B

Diaphragm

Arm lot

Fig. 10-4 Exploded view of diaphragm type aspirator

– 31 –

Image 39
Contents Infrared GAS Analyzer Request IntroductionProhibition Iii Prohibition Others Contents Vii Prepared SpecificationsOutline CompositionInstallation Selection of installation siteSelection of gas sampling location Connection of cables Installation workPiping procedure Preparation for Operation Injection of clean waterSupply of external power Description of Each PartSetting of temperature inside locker Airtightness testWarm-up time SO2 INDICATION/OPERATION PanelThis panel consists of the following functions Function and indication of indication/operation panelGeneral operation of indication/operation panel Advice on OperationCalibration Kinds of standard gasPreparation of standard gas Pressure regulation for standard gas cylinder Func CAL SET ENT Calibration concentration setting except zirconia O2 sensorCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operation Operating procedureRoutine maintenance and inspection MaintenanceConditioner 200ppm or more at 0.5L/min Periodic maintenance and inspectionOther maintenance and inspection Centration is measured in the setting range of 0 toGas extractor type ZBAK1 Maintenance of sampling deviceMembrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Others, major component figures Maintenance of zirconia O2 sensorStandard accessories Standard Accessories and SparesCode for preparation of 1-year spares and supplies Spares for 1 year optionalOutline Processing of Heating Pipe EndsEnd processing method Grid 300mm 40mm Grid area 150 mm Clamp support Check after pipingSpecifications of heating pipe Specifications of tube traceTroubleshooting of instrument TroubleshootingError codes and remedies Mass weight Ambient temperaturePower supply Power consumptionStandard requirements For sample gas Month/NO2, 100ppmNumber of output points and kinds of outputs Moving average output Block diagramCalibration cycle Once/day variable, up to 7 days Automatic calibrationOrdering information Calibration tolerance Automatic calibration output holdCode Symbols Diaglam Functions of individual components For indoor installation type For outdoor installation typeOnly specified outputs are wired Connection DiagramHead Office Instrumentation Div International Sales Dept