Page 23

Setting of span calibration concentration (zero point stipulated in the laws for measurement and weigh- ing)

Press the CAL key in the status where the digit of the O2 analyzer main indication is flickering via the zero (air) point calibration concentration setting.

The previously set span value will now appear on the main indication.

The ‘CAL SET LED’ of the function indicating lamps will flicker.

Then, press the COMP key and the highest digit of the 02 analyzer main indication will flicker and the span value is now settable.

Press the keys to select a range.

When the range has been selected, set a span value in this status.

The numeric value will be incremented by pressing

the key.

Press the key and the digit to be set can be se- lected.

Press the ENT key after setting the span value. Set- ting operation is completed.

Setting of each component

O2

CAL

NO

COMP

O2

O2

ENT

O2

– 15 –

Image 23
Contents Infrared GAS Analyzer Request IntroductionProhibition Iii Prohibition Others Contents Vii Prepared SpecificationsOutline CompositionSelection of gas sampling location InstallationSelection of installation site Connection of cables Installation workPiping procedure Preparation for Operation Injection of clean waterSupply of external power Description of Each PartWarm-up time Setting of temperature inside lockerAirtightness test SO2 INDICATION/OPERATION PanelThis panel consists of the following functions Function and indication of indication/operation panelGeneral operation of indication/operation panel Advice on OperationPreparation of standard gas CalibrationKinds of standard gas Pressure regulation for standard gas cylinder Func CAL SET ENT Calibration concentration setting except zirconia O2 sensorCAL Calibration concentration setting zirconia O2 sensor CAL Comp ENT Zero Single-component analyzer Auto calibration Auto CAL Func Auto CAL Func Auto CAL ENT Auto calibration ON/OFF setting Func Hold ENT Operation Operating procedureRoutine maintenance and inspection MaintenanceConditioner 200ppm or more at 0.5L/min Periodic maintenance and inspectionOther maintenance and inspection Centration is measured in the setting range of 0 toGas extractor type ZBAK1 Maintenance of sampling deviceMembrane filter Exploded view of diaphragm type aspirator Structure of electronic dehumidifier Structure of NO2 /NO converter Structure of Permapure dryer Others, major component figures Maintenance of zirconia O2 sensorStandard accessories Standard Accessories and SparesCode for preparation of 1-year spares and supplies Spares for 1 year optionalOutline Processing of Heating Pipe EndsEnd processing method Grid 300mm 40mm Grid area 150 mm Clamp support Check after pipingSpecifications of heating pipe Specifications of tube traceTroubleshooting of instrument TroubleshootingError codes and remedies Mass weight Ambient temperaturePower supply Power consumptionNumber of output points and kinds of outputs Standard requirements For sample gasMonth/NO2, 100ppm Moving average output Block diagramCalibration cycle Once/day variable, up to 7 days Automatic calibrationOrdering information Calibration tolerance Automatic calibration output holdCode Symbols Diaglam Functions of individual components For indoor installation type For outdoor installation typeOnly specified outputs are wired Connection DiagramHead Office Instrumentation Div International Sales Dept