Chapter 4: Installing Oscilloscope Measurement Modules

16533/34A Oscilloscope Module (single or multi-card modules)

 

 

16533/34A Oscilloscope Module (single or multi-card

 

 

modules)

 

 

The Agilent Technologies 16533A/34A oscilloscope module functions as either

 

 

a master card or expander card. It is compatible with all versions of software

 

 

in your 16700A/B or 16702A/B mainframe. The circuitry in the module

 

 

requires an operational accuracy calibration to optimize measurement

 

 

accuracy.

 

 

A multicard module should contain either all 16533A or 16534A cards.

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTE:

 

Each of the individual cards of a multicard 16533A or a multicard 16534A

 

 

module must first be calibrated as a single card. After reconfiguring into a

 

 

multicard module, the channel skew calibration needs to be done.

 

 

Step 1

 

 

 

 

Prepare a single 16533/34A card for calibration

 

 

 

 

 

 

 

 

 

 

 

 

 

CAUTION:

 

The effects of ELECTROSTATIC DISCHARGE can damage components. Use

 

 

grounded wrist straps and mats when you are performing any kind of service

 

 

on this module.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1 Power down your 16700A/B or 16702A/B mainframe.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16700A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16700B or 16702B

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2Remove all modules and filler panels from the mainframe and set the PROTECTED/UNPROTECTED switch to UNPROTECTED.

72

Page 72
Image 72
Agilent Technologies 16700B manual Modules, Prepare a single 16533/34A card for calibration