Chapter 4: Installing Oscilloscope Measurement Modules
16533/34A Oscilloscope Module (single or
|
| 16533/34A Oscilloscope Module (single or | ||||||||||
|
| modules) | ||||||||||
|
| The Agilent Technologies 16533A/34A oscilloscope module functions as either | ||||||||||
|
| a master card or expander card. It is compatible with all versions of software | ||||||||||
|
| in your 16700A/B or 16702A/B mainframe. The circuitry in the module | ||||||||||
|
| requires an operational accuracy calibration to optimize measurement | ||||||||||
|
| accuracy. | ||||||||||
|
| A multicard module should contain either all 16533A or 16534A cards. | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
NOTE: |
| Each of the individual cards of a multicard 16533A or a multicard 16534A | ||||||||||
|
| module must first be calibrated as a single card. After reconfiguring into a | ||||||||||
|
| multicard module, the channel skew calibration needs to be done. | ||||||||||
|
| Step 1 | ||||||||||
|
| |||||||||||
|
| Prepare a single 16533/34A card for calibration | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
CAUTION: |
| The effects of ELECTROSTATIC DISCHARGE can damage components. Use | ||||||||||
|
| grounded wrist straps and mats when you are performing any kind of service | ||||||||||
|
| on this module. | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| 1 Power down your 16700A/B or 16702A/B mainframe. | |||||||||||
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
| 16700A |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 16700B or 16702B | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
2Remove all modules and filler panels from the mainframe and set the PROTECTED/UNPROTECTED switch to UNPROTECTED.
72