38
Chapter 2: Connecting and Configuring Hardware
To connect multiple frames
To connect multiple frames
As many as eight 16700B’s and/or 16702B’s with expander frames may be
connected together. To connect multiple frames you need to order 16700B
option #012 and/or 16702B option #012.
NOTE: The multiframe module requires software Rev. A.02.00.00 or higher. Agilent
16700B and 16702B logic analysis systems ordered with the multiframe option
installed will have the current operating system software ins talled.
1If the multiframe module is already installed, skip to step 11.
2End your logic analysis session.
aExit all logic analysis sessions. In the session manager, select
Shutdown.
bAt the query, select Powerdown.
cWhen the “OK to turn off power or reset system” message appears,
turn the instrument off.
dRemove power from the instrument.
3Disconnect the power cable and all data and peripheral cables from the
rear panel.
4Move the instrument to a static-safe work area.
5Lay the instrument on its side so the handle side is up.
6Using a Torx T10 screwdriver, remove two feet and one screw at the
center rear of the cover that secures the bottom cover to the frame.