Chapter 2: Connecting and Configuring Hardware
To connect multiple frames
| To connect multiple frames |
| As many as eight 16700B’s and/or 16702B’s with expander frames may be |
| connected together. To connect multiple frames you need to order 16700B |
| option #012 and/or 16702B option #012. |
|
|
NOTE: | The multiframe module requires software Rev. A.02.00.00 or higher. Agilent |
| 16700B and 16702B logic analysis systems ordered with the multiframe option |
| installed will have the current operating system software installed. |
|
|
1If the multiframe module is already installed, skip to step 11.
2End your logic analysis session.
a Exit all logic analysis sessions. In the session manager, select Shutdown.
b At the query, select Powerdown.
c When the “OK to turn off power or reset system” message appears, turn the instrument off.
d Remove power from the instrument.
3Disconnect the power cable and all data and peripheral cables from the rear panel.
4Move the instrument to a
5Lay the instrument on its side so the handle side is up.
6Using a Torx T10 screwdriver, remove two feet and one screw at the center rear of the cover that secures the bottom cover to the frame.
38