Chapter 4: Installing Oscilloscope Measurement Modules

16533/34A Oscilloscope Module (single or multi-card modules)

3Select the Run icon and the instrument will remind you to connect the cables. Select OK and wait for the trigger skew calibration to complete.

RUN

As trigger skew calibration runs, messages appear in the message box on screen. When the Ext Trig Skew calibration is complete, the resulting calibration factors are stored in non-volatile RAM.

4If a multi-card module is being calibrated, repeat the procedures beginning with Step 1 on page 72 for each card until all cards have been calibrated.

5If all cards have been calibrated, remove the BNC cables from the instrument.

6Select Close in the Calibration window.