Chapter 4: Installing Oscilloscope Measurement Modules

16533/34A Oscilloscope Module (single or multi-card modules)

b Repeat for all cards in the module. Up to 4 cards may be configured on￿￿￿￿￿a single$ time base and trigger in a 16700A, 16700B, 16702A, or 16702B mainframe

2 Reapply power to the 16700-series mainframe.

3 In the Logic Analysis System window select the icon for the master card of the multi-card module then select Calibration.

4Perform channel skew calibration on the multi-card module.

a Connect two 9-inch 50-ohm BNC cables and a BNC tee adapter between channel 1, AC/DC cal, of the first card and channel 1 of the second card.

Chan. 1

Chan. 2

Chan. 1

Chan. 2

 

 

 

 

Channel 1 Upper Card

AC/CD Cal

 

Upper Card

Channel 1

 

Lower Card

82

Page 82
Image 82