81
Chapter 4: Installing Oscilloscope Measurement Modules
16533/34A Oscilloscope Module (single or multi-card modules)
10 If you are only calibrating one card set the PROTECTED/
UNPROTECTED switch back to PROTECTED.
11 For multi-card modules, leave the switch set to UNPROTECTED and
continue to Step 4 to perform reconfiguration.
Step 4Reconfigure multi-card modules
A multicard module should contain either all 16533A or 16534A cards.
1After calibrating each card individually (following the procedures
beginning on page 72) and before applying power to the mainframe,
connect the module cables.
aBeginning with the top-most card, connect the ECL EXT TRIG OUT
to the ECL EXT TRIG IN of the card immediately below. Use the
master/expander trigger cable included with the accessory kit of
each card.
PROTECTED
UNPROTECTED
If only calibrating one
card set protect.
Master/expander
trigger cables