Chapter 4 Calibration Procedures
Frequency Verification
Frequency Verification
This test verifies the frequency accuracy of the two sources in the function generator. All output frequencies are derived from a single generated frequency, and only one frequency point is checked.
The second test verifies the burst rate frequency.
Set the function generator for each output indicated in the table below. Use a frequency meter to measure the output frequency. Compare the measured results to the test limits shown in the table. This is a 50W output termination test.
|
|
|
|
| Agilent 33120A |
|
| Measurement | |||
|
|
| Function | OUT 1 | Ampl | Freq | BURST | BURST | Nominal | Error | |
|
|
| TERM | RATE | CNT | ||||||
|
|
|
|
|
|
|
|
| |||
Q | Sine wave | 50 W | 3.5 Vrms | 1.00 kHz | — | — | 1.00 kHz | – 0.02 Hz | |||
Q | Square wave | 50 W | 3.5 Vrms | 1.00 kHz | 500 Hz | 1 CYC | 500 Hz | – 5 Hz | |||
|
|
|
|
|
|
|
|
|
|
|
|
Function Gain and Linearity Verification
Q
Q
1
This test verifies the output amplitude accuracy specification for sine wave, triangle wave, ramp, and square wave outputs.
Set the function generator for each output indicated in the table below. Use a DMM to measure the function generator ACrms output voltage. Compare the measured results to the test limits shown in the table.
This is a HIGH Z output termination test.
| Agilent 33120A |
| Measurement | ||
Function | OUT TERM 1 | Ampl | Freq | Nominal | Error |
Sine wave | HIGH Z | 7.0 Vrms | 1.0 kHz | 7.0 Vrms | – 0.07 Vrms |
|
|
|
|
|
|
Sine wave | HIGH Z | 5.7 Vrms | 1.0 kHz | 5.7 Vrms | – 0.057 Vrms |
|
|
|
|
|
|
Triangle wave | HIGH Z | 5.7 Vrms | 100 Hz | 5.7 Vrms | – 0.057 Vrms |
Ramp wave | HIGH Z | 5.7 Vrms | 100 Hz | 5.7 Vrms | – 0.057 Vrms |
|
|
|
|
|
|
Square wave | HIGH Z | 10.0 Vrms | 100 Hz | 10.0 Vrms | – 0.1 Vrms |
|
|
|
|
|
|
Square wave | HIGH Z | 8.0 Vrms | 100 Hz | 8.0 Vrms | – 0.08 Vrms |
|
|
|
|
|
|
Output termination set using front panel controls. HIGH Z assumes no load on output. 50W assumes a 50W – 0.1W load on output.
56