Chapter 4 Calibration Procedures
Calibration Security Code
To Unsecure the Function Generator Without the Security Code
To unsecure the function generator without the correct security code, follow the steps below. A procedure to unsecure the function generator is given on page 47. Also see “Electrostatic Discharge (ESD) Precautions” in chapter 6 before beginning this procedure.
SHOCK HAZARD. Only
W A R N I N G of the hazards involved should remove the instrument covers. The procedures in this section require that you connect the power cord to the instrument with the covers removed. To avoid electrical shock and personal injury, be careful not to touch the
1Disconnect the power cord and all input connections (front and rear terminals).
2Remove the instrument cover. Refer to the disassembly drawing on page 130.
3 Connect the power cord and turn on the function generator. | 4 |
|
4Apply a short between the two exposed metal pads on JM101 (located near U106 and U205) as shown in the figure below.
5While maintaining the short, enter any unsecure code. The function generator is now unsecured.
6Remove the short at JM101.
7Turn off and reassemble the function generator.
8The function generator is now unsecured and you can enter a new security code. Be sure you take note of the new security code.
65