Chapter 4 Calibration Procedures
General Calibration/Adjustment Procedure
General Calibration/Adjustment Procedure
The adjustment procedures described in chapter 4 use the CAL MENU to generate and set internal calibration constants. The general menu procedure is the same for all calibration setups. The following example demonstrates making the Frequency and Burst Rate adjustments.
1 Turn on the menu.
A: MOD MENU
2Move across to the CAL MENU choice on this level. 1
F: CAL MENU
3 Move down a level to the UNSECURED command.
1: UNSECURED
4
>
Enter
If the display shows SECURED, you will have to unsecure the function generator to perform the calibration procedures. A procedure is given on page 47.
4Move across to the CALIBRATE choice. 1
2: CALIBRATE
5Move down one level.
The display indicates the calibration setup number. You can change this number to perform individual specification adjustments.
SETUP 00
6 Begin the Frequency and Burst Rate adjustment procedure.