Chapter 4 Calibration Procedures

General Calibration/Adjustment Procedure

General Calibration/Adjustment Procedure

The adjustment procedures described in chapter 4 use the CAL MENU to generate and set internal calibration constants. The general menu procedure is the same for all calibration setups. The following example demonstrates making the Frequency and Burst Rate adjustments.

1 Turn on the menu.

A: MOD MENU

2Move across to the CAL MENU choice on this level. 1

F: CAL MENU

￿￿

3 Move down a level to the UNSECURED command.

1: UNSECURED

4

>

￿￿

Enter

If the display shows SECURED, you will have to unsecure the function generator to perform the calibration procedures. A procedure is given on page 47.

4Move across to the CALIBRATE choice. 1

2: CALIBRATE

5Move down one level.

The display indicates the calibration setup number. You can change this number to perform individual specification adjustments.

SETUP 00

6 Begin the Frequency and Burst Rate adjustment procedure.

67

Page 69
Image 69
Agilent Technologies 33120A manual General Calibration/Adjustment Procedure, Move down a level to the Unsecured command