Chapter 6 Service

To Replace the Power-Line Fuse

To Replace the Power-Line Fuse

The power-line fuse is located within the function generator’s fuse-holder assembly on the rear panel (see page 22). The function generator is shipped from the factory with a 500 mAT slow-blow fuse installed (part number 2110-0458). This is the correct fuse for all line voltages.

To Replace the Output Protection Fuse (F801)

The Output Protection Fuse is located inside the function generator. This fuse is a thru-hole soldered 500 mA part (part number 2110-0716). The fuse is located near the output connector (J801) on the main PC board. You will need to disassemble the function generator to replace this fuse (use a TORX T-15 driver to remove the screws located on the rear panel). The disassembly procedure is shown on page 130.

6

107

Page 109
Image 109
Agilent Technologies 33120A manual To Replace the Power-Line Fuse, To Replace the Output Protection Fuse F801