Chapter 4 Calibration Procedures

AC Amplitude Adjustment (50W)

AC Amplitude Adjustment (50W)

1The function generator stores 16 calibration constants related to 50W output. The constants are calculated from the adjustment value entered. The calibration constants are stored following completion of setup 49 and the calibration procedure may be aborted after that point. No calibration constants are stored if the procedures are aborted at any other setup.

2Use the DMM to measure the resistance of a 50W feedthrough load. Record the measurement for step 3. You can measure the load and cable resistance (recommended procedure) or just the load as shown below.

4

3Enter the following setup and use the calibrate menu to enter the measured value of the 50W feedthrough load (and cable). This number will be used to calculate the 50W output amplitude calibration constants.

 

Nominal Input

 

 

 

 

SETUP

LOAD Z

 

33

50 W

Enter measured value of load.

 

 

 

Once the value of the 50W load and cable are entered, use the SAME load and cable for all 50W tests.

73