Shift
Menu On/Off
<
Chapter 3
To unsecure the function generator for calibration
To unsecure the function generator for calibration
The function generator can use a calibration security code to prevent unauthorized or accidental calibration. This procedure shows you how to unsecure the function generator for calibration.
1Turn on the menu.
A: MOD MENU
3
2Move across to the CAL MENU choice on this level.
F:CAL MENU
3Move down a level to the SECURED command.
1: SECURED
If the display shows UNSECURED, you do not need to perform this procedure to execute a calibration.