Chapter 3 Front-Panel Menu Operation

To unsecure the function generator for calibration

To unsecure the function generator for calibration

The function generator can use a calibration security code to prevent unauthorized or accidental calibration. This procedure shows you how to unsecure the function generator for calibration.

1Turn on the menu.

A: MOD MENU

3

2Move across to the CAL MENU choice on this level.

F:CAL MENU

3Move down a level to the SECURED command.

1: SECURED

If the display shows UNSECURED, you do not need to perform this procedure to execute a calibration.

47

Page 49
Image 49
Agilent Technologies 33120A To unsecure the function generator for calibration, Move down a level to the Secured command