Nominal Output
SETUP FREQUENCY AMPLITUDE Adjustment for:
8 1 kHz 5.5 V rms 2 dB Output Attenuator
9 1 kHz 4.4 V rms 4 dB Output Attenuator
10 1 kHz 3.5 V rms 6 dB Output Attenuator
11 1 kHz 2.8 V rms 8 dB Output Attenuator
12 1 kHz 2.2 V rms 10 dB Output Attenuator
13 1 kHz 1.7 V rms 12 dB Output Attenuator
14 1 kHz 1.4 V rms 14 dB Output Attenuator
15 1 kHz 1.1 V rms 16 dB Output Attenuator
16 1 kHz 0.88 V rms 18 dB Output Attenuator
17 1 kHz 0.70 V rms 20 dB Output Attenuator
18 1 kHz 0.55 V rms 22 dB Output Attenuator
19 1 kHz 0.44 V rms 24 dB Output Attenuator
20 1 kHz 0.35 V rms 26 dB Output Attenuator
21 1 kHz 0.28 V rms 28 dB Output Attenuator
22 1 kHz 0.22 V rms 30 dB Output Attenuator
23 1 kHz 5.5 V rms 2 dB Pre-attenuator
24 1 kHz 4.4 V rms 4 dB Pre-attenuator
25 1 kHz 3.5 V rms 6 dB Pre-attenuator
26 1 kHz 2.8 V rms 8 dB Pre-attenuator
27 1 kHz 2.2V rm s 10 dB Pre-attenuator
28 1 kHz 1.7 V rms 12 dB Pre-attenuator
29 1 kHz 1.4 Vrm s 14 dB Pre-attenuator
2Use the CALIBRATE menu to adjust the displayed amplitude at each
setup to match the measured amplitude and enter the value.
3Perform the AC Amplitude Verification procedures on page 57.
4
Chapter 4 Calibration Procedures
AC Amplitude Adjustment (High-Z)
71