Chapter 4 Calibration Procedures
DC Output Adjustment
DC Output Adjustment
The function generator stores nine calibration constants related to
DC volts output. The constants are calculated from the adjustment value entered. The calibration constants are stored following completion of setup 59. No calibration constants are stored if the procedures are aborted at any other setup.
1Use a DMM to measure the function generator dcV output voltage for each setup in the following table. These adjustments use a HIGH Z output termination.
| Nominal Output |
|
|
|
|
SETUP | DC Volts | Adjustment for: |
|
|
|
50 | - 8.0 Vdc | Negative offset gain |
51 | 8.0 Vdc | Positive offset gain |
|
|
|
52 | 0.0 Vdc | AM offset |
53 | 0.0 Vdc | 2 dB |
|
|
|
54 | 0.0 Vdc | 4 dB |
55 | 0.0 Vdc | 6 dB |
|
|
|
56 | 0.0 Vdc | 8 dB |
|
|
|
57 | 0.0 Vdc | 10 dB |
|
|
|
58 | 0.0 Vdc | 12 dB |
|
|
|
59 | 0.0 Vdc | 14 dB |
|
|
|
2Use the CALIBRATE menu to adjust the displayed output voltage at each setup to match the measured voltage and enter the value.
3Perform the DC Function Offset Verification procedures on page 57.
76