Chapter 4 Calibration Procedures

Duty Cycle Adjustment

Duty Cycle Adjustment

The function generator stores two calibration constants related to squarewave offset and two calibration constants related to squarewave duty cycle. The constants are calculated from the adjustment value entered. The calibration constants are stored following completion of setup 63. No calibration constants are stored if the procedures are aborted at any other setup.

1Use a DMM to measure the function generator dcV output voltage for each setup in the following table. These adjustments use a HIGH Z output termination.

For this test, the DMM must be set to a fixed range capable of measuring from +10 V to -10 V. Do not use an auto-ranging function for this test.

4

 

Nominal Output

 

 

 

 

 

SETUP

FREQUENCY

AMPLITUDE

 

60

10.0 Vdc

Positive squarewave offset.

 

 

 

 

61

-10.0 Vdc

Negative squarewave offset.

 

 

 

 

62

300 Hz

0.0 Vdc

50% duty cycle squarewave.

 

 

 

 

63

300 Hz

5.0 Vdc

75% duty cycle squarewave

 

 

 

 

2Use the CALIBRATE menu to adjust the displayed output voltage at each setup to match the measured voltage and enter the value.

3Perform the Squarewave Duty Cycle Verification procedures on page 62.

AC Amplitude Flatness Adjustment

The function generator stores eleven calibration constants related to AC Amplitude Flatness from 1 kHz to 15 MHz. The constants are calculated from the adjustment value entered and one of two calculation constants related to the type of measurement device you are using. The calibration constants are stored following completion of setup 82. No calibration constants are stored if the procedures are aborted at any other setup.

77

Page 79
Image 79
Agilent Technologies 33120A manual Duty Cycle Adjustment, AC Amplitude Flatness Adjustment