Agilent 33120A Measurement
Function OUT TERM 1Ampl Freq Nominal Error
QSine wave HIGH Z 7.0 Vrms 1.00 kHz 7.0 Vrms 0.070 Vrms
Sine wave HIGH Z 5.7 Vrms 1.00 kHz 5.7 Vrms 0.057 Vrms
Sine wave HIGH Z 5.5 Vrms 1.00 kHz 5.5 Vrms 0.055 Vrms
Sine wave HIGH Z 4.4 Vrms 1.00 kHz 4.4 Vrms 0.044 Vrms
Sine wave HIGH Z 3.5 Vrms 1.00 kHz 3.5 Vrms 0.035 Vrms
Sine wave HIGH Z 2.8 Vrms 1.00 kHz 2.8 Vrms 0.028 Vrms
Sine wave HIGH Z 2.2 Vrms 1.00 kHz 2.2 Vrms 0.022 Vrms
Sine wave HIGH Z 1.7 Vrms 1.00 kHz 1.7 Vrms 0.017 Vrms
Sine wave HIGH Z 1.4 Vrms 1.00 kHz 1.4Vrms 0.014 Vrms
Sine wave HIGH Z 1.1 Vrms 1.00 kHz 1.1 Vrms 0.011 Vrms
QSine wave HIGH Z 0.88 Vrms 1.00 kHz 0.88 Vrms 0.0088 Vrms
Sine wave HIGH Z 0.70 Vrms 1.00 kHz 0.70 Vrms 0.0070 Vrms
Sine wave HIGH Z 0.55 Vrms 1.00 kHz 0.55 Vrms 0.0055 Vrms
Sine wave HIGH Z 0.44 Vrms 1.00 kHz 0.44 Vrms 0.0044 Vrms
Sine wave HIGH Z 0.35 Vrms 1.00 kHz 0.35 Vrms 0.0035 Vrms
Sine wave HIGH Z 0.28 Vrms 1.00 kHz 0.28 Vrms 0.0028 Vrms
Sine wave HIGH Z 0.22 Vrms 1.00 kHz 0.22 Vrms 0.0022 Vrms
Sine wave HIGH Z 0.17 Vrms 1.00 kHz 0.17 Vrms 0.0017 Vrms
Sine wave HIGH Z 0.14 Vrms 1.00 kHz 0.14 Vrms 0.0014 Vrms
Sine wave HIGH Z 0.11 Vrms 1.00 kHz 0.11 Vrms 0.0011 Vrms
QSine wave HIGH Z 0.088 Vrms 1.00 kHz 0.088Vrms 0.00088 Vrms
Sine wave HIGH Z 0.070 Vrms 1.00 kHz 0.070 Vrms 0.00070 Vrms
Sine wave HIGH Z 0.055 Vrms 1.00 kHz 0.055 Vrms 0.00055 Vrms
Sine wave HIGH Z 0.044 Vrms 1.00 kHz 0.044 Vrms 0.00044 Vrms
QSine wave HIGH Z 0.036 Vrms 1.00 kHz 0.036 Vrms 0.00036 Vrms
1Output termination set using front panel controls. HIGH Z assumes no load on
output. 50W assumes a 50W 0.1W load on output.
Chapter 4 Calibration ProceduresAC Amplitude Verification 58