Chapter 4 Calibration Procedures
DC Function Offset Verification
DC Function Offset Verification
This test verifies the DC offset and DC output specifications.
Set the function generator for each output indicated in the table below. Use a DMM to measure the function generator dcV output. Compare the measured results to the test limits shown in the table. This is a HIGH Z output termination test.
|
| Agilent 33120A |
| Measurement | |
|
|
|
|
|
|
| Function | OUT TERM 1 | Ampl | Nominal | Error |
Q | DC Volts | HIGH Z | 10.0 Vdc | 10.0 Vdc | – 0.20 Vdc |
|
|
|
|
|
|
| DC Volts | HIGH Z | – 0.20 Vdc | ||
|
|
|
|
|
|
4
AC Amplitude Verification
This procedure is used to check the output amplitude calibration of the function generator. Verification checks are performed to check the accuracy of the
Set the function generator for each output indicated in the table on the next page. Use a DMM to measure the ACrms output voltage of the function generator. Compare the measured results to the test limits shown in the table. This is a HIGH Z output termination test.
1Output termination set using front panel controls. HIGH Z assumes no load on output. 50W assumes a 50W