Chapter 4 Calibration Procedures

AM Modulation Depth Verification

AM Modulation Depth Verification

This test verifies the modulation depth specification.

Select each function generator output in the table below. Use a DMM to measure the function generator ACrms output voltage. Compare the measured results to the test limits shown in the table. This is a HIGH Z output termination test.

 

 

 

Agilent 33120A

 

 

 

 

Measurement

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AM Modulation

 

 

 

 

Function

OUT 1

Ampl

Freq

 

Shape

Freq

 

Depth

Nominal

Error

 

TERM

 

 

 

 

 

 

 

 

 

 

 

 

 

Q

Sine wave

HIGH Z

1.0 Vrms

1.00 kHz

 

Sinewave

100 Hz

 

0%

0.50 Vrms 0.005 Vrms

 

Sine wave

HIGH Z

1.0 Vrms

1.00 kHz

 

Sinewave

100 Hz

 

100%

0.61 Vrms 0.0061 Vrms

 

 

 

 

 

 

 

 

 

 

 

 

4

1Output termination set using front panel controls. HIGH Z assumes no load on output. 50W assumes a 50W – 0.1W load on output.

61