Chapter 4 Calibration Procedures
AM Modulation Depth Verification
AM Modulation Depth Verification
This test verifies the modulation depth specification.
Select each function generator output in the table below. Use a DMM to measure the function generator ACrms output voltage. Compare the measured results to the test limits shown in the table. This is a HIGH Z output termination test.
|
|
| Agilent 33120A |
|
|
|
| Measurement | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| AM Modulation |
|
|
| ||
| Function | OUT 1 | Ampl | Freq |
| Shape | Freq |
| Depth | Nominal | Error |
| TERM |
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
| |
Q | Sine wave | HIGH Z | 1.0 Vrms | 1.00 kHz |
| Sinewave | 100 Hz |
| 0% | 0.50 Vrms – 0.005 Vrms | |
| Sine wave | HIGH Z | 1.0 Vrms | 1.00 kHz |
| Sinewave | 100 Hz |
| 100% | 0.61 Vrms – 0.0061 Vrms | |
|
|
|
|
|
|
|
|
|
|
|
|
4
1Output termination set using front panel controls. HIGH Z assumes no load on output. 50W assumes a 50W – 0.1W load on output.