Chapter 4 Calibration Procedures

General Calibration/Adjustment Procedure

7Move the flashing cursor over the digit to be edited. 1

8Change the value in the display to match the measured frequency.

1.000,0040KHz

9Calculate and save the new value.

CALIBRATING

10Perform the next adjustment procedure.

The setup number and function generator output is automatically set for the next adjustment procedure.

SETUP ^01

You will press ENTER twice for each calibration step, once to select the setup (as described in step 6) and once to enter the adjustment (as described in step 9).

1You can also use the knob to scroll left or right through the choices on each level of the menu

68

Page 70
Image 70