Chapter 4 Calibration Procedures

Optional Performance Verification Tests

Distortion Verification

This test checks the Harmonic Distortion at selected frequencies and harmonics. This test requires the use of a spectrum analyzer with dynamic range, frequency range, and resolution bandwidth adequate for the measurement.

Select each function generator output in the table below. Use a spectrum analyzer connected to the function generator output. Set the fundamental frequency reference to 0 dB and measure the 2nd through 5th harmonic frequencies relative to this reference. This test is a 50W output termination test.

 

Agilent 33120A

 

 

 

Measurement

 

 

 

 

 

 

 

 

 

 

harmonic

 

 

 

 

Function

OUT 1

Ampl

Freq

Fundamental

2nd

3rd

4th

5th

Value below

 

 

TERM

reference

 

 

 

 

 

 

 

 

 

 

 

4

 

 

 

 

 

 

 

 

 

 

 

Sine wave

50 W

1.1 Vrms

20.00 kHz

reference

40 kHz

60 kHz

80 kHz

100 kHz

< 70 dB

Sine wave

50 W

1.1 Vrms

100.00 kHz

reference

200 kHz

300 kHz

400 kHz

500 kHz

< 60 dB

 

 

 

 

Sine wave

50 W

1.1 Vrms

1.00 MHz

reference

2 MHz

3 MHz

4 MHz

5 MHz

< 45 dB

 

 

Sine wave

50 W

1.1 Vrms

15.00 MHz

reference

30 MHz

45 MHz

60 MHz

75 MHz

< 35 dB

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1Output termination set using front panel controls. HIGH Z assumes no load on output. 50W assumes a 50W – 0.1W load on output.