Chapter 1 Specifications

Agilent 33120A Function Generator

1

MODULATION CHARACTERISTICS

AM Modulation

 

Carrier -3 dB Freq:

10 MHz (typical)

Modulation:

Any internal waveform plus Arb

Frequency:

10 mHz to 20 kHz (￿￿￿0.05% to

 

2.5 kHz, then decreases linearly

 

to ￿￿0.4% at upper limit)

Depth:

0% to 120%

Source:

Internal / External

FM Modulation

 

Modulation:

Any internal waveform plus Arb

Frequency:

10 mHz to 10 kHz (￿￿￿0.05% to

 

600 Hz, then decreases linearly

 

to ￿￿0.8% at upper limit)

Peak Deviation:

10 mHz to 15 MHz

Source:

Internal Only

Burst Modulation

5 MHz max.

Carrier Frequency:

Count:

1 to 50,000 cycles, or Infinite

Start Phase:

-360￿to +360￿

Internal Rate:

10 mHz to 50 kHz ￿￿1%

Gate Source:

Internal or External Gate (1)

Trigger Source:

Single, External, or Internal Rate

FSK Modulation

10 mHz to 15 MHz (￿￿￿0.05% to

Frequency Range:

 

600 Hz, then decreases linearly

 

to ￿￿4% at upper limit)

Internal Rate:

10 mHz to 50 kHz

Source:

Internal / External (1 MHz max.)

FREQUENCY SWEEP

Type:

Linear or Logarithmic

Direction:

Up or Down

Start F / Stop F:

10 mHz to 15 MHz

Time:

1 ms to 500 sec ￿￿0.1%

Source:

Single, External, or Internal

REAR-PANEL INPUTS

External AM

￿5 Vpk = 100% Modulation

Modulation:

5 k￿Input Resistance

External Trigger/FSK

 

Burst Gate: (1)

TTL (low true)

Latency:

1.3 ￿s

Jitter:

25 ns

SYSTEM CHARACTERISTICS

 

Configuration Times (2)

 

 

 

 

 

 

Function Change: (3)

 

 

80 ms

 

Frequency Change: (3)

 

 

30 ms

 

Amplitude Change:

 

 

30 ms

 

Offset Change:

 

 

10 ms

 

Select User Arb:

 

 

100 ms

 

Modulation Parameter Change:

< 350 ms

Arb Download Times over GPIB:

 

 

 

 

Arb Length

 

Binary

 

ASCII Integer

 

ASCII Real (4)

 

 

 

 

 

 

 

16,000 points

 

8 sec

 

81 sec

 

 

100 sec

 

8,192 points

 

4 sec

 

42 sec

 

 

51 sec

 

4,096 points

 

2.5 sec

 

21 sec

 

 

26 sec

 

2,048 points

 

1.5 sec

 

11 sec

 

 

13 sec

Arb Download Times over RS-232 at 9600 Baud: (5)

Arb Length

 

Binary

 

ASCII Integer

 

ASCII Real (6)

 

 

 

 

16,000 points

 

35 sec

 

101 sec

 

134 sec

8,192 points

 

18 sec

 

52 sec

 

69 sec

4,096 points

 

10 sec

 

27 sec

 

35 sec

2,048 points

 

6 sec

 

14 sec

 

18 sec

(1)Trigger source ignored when External Gate is selected.

(2)Time to change parameter and output the new signal.

(3)Modulation or sweep off.

(4)Times for 5-digit and 12-digit numbers.

(5)For 4800 baud, multiply the download times by two;

For 2400 baud, multiply the download times by four, etc.

(6)Time for 5-digit numbers. For 12-digit numbers, multiply the 5-digit numbers by two.