Agilent Technologies 33120A manual Adjustment for KHz

Models: 33120A

1 146
Download 146 pages 51 Kb
Page 77
Image 77

Chapter 4 Calibration Procedures

AC Amplitude Adjustment (50W)

 

Nominal Output

 

 

 

SETUP

FREQUENCY

AMPLITUDE

Adjustment for:

 

 

 

 

 

 

 

 

34

1 kHz

3.5 Vrms

0 dB Output Attenuator

 

 

35

1 kHz

2.8 Vrms

2 dB Output Attenuator

 

 

 

 

 

 

 

 

36

1 kHz

2.23 Vrms

4 dB Output Attenuator

 

 

 

 

 

 

 

 

37

1 kHz

1.77 Vrms

6 dB Output Attenuator

 

 

 

 

 

 

 

 

38

1 kHz

1.41 Vrms

8 dB Output Attenuator

 

 

 

 

 

 

 

 

39

1 kHz

1.12 Vrms

10 dB Output Attenuator

 

 

 

 

 

 

 

 

40

1 kHz

.887 Vrms

12 dB Output Attenuator

 

 

 

 

 

 

 

 

41

1 kHz

.704 Vrms

14 dB Output Attenuator

 

 

42

1 kHz

.559 Vrms

16 dB Output Attenuator

 

 

43

1 kHz

.442 Vrms

18 dB Output Attenuator

 

 

44

1 kHz

.350 Vrms

20 dB Output Attenuator

 

 

 

 

 

 

 

 

44

1 kHz

.281 Vrms

22 dB Output Attenuator

 

4

 

 

 

 

 

46

1 kHz

.223 Vrms

24 dB Output Attenuator

 

 

47

1 kHz

.177 Vrms

26 dB Output Attenuator

 

 

 

 

 

 

 

 

48

1 kHz

.141 Vrms

28 dB Output Attenuator

 

 

49

1 kHz

.112 Vrms

30 dB Output Attenuator

 

 

 

 

 

 

 

 

5Use the CALIBRATE menu to adjust the displayed amplitude at each setup to match the measured amplitude and enter the value.

6Perform the AC Amplitude Verification procedures beginning on page 57.