Agilent Technologies 33120A manual AC Amplitude Flatness Adjustment

Models: 33120A

1 146
Download 146 pages 51 Kb
Page 80
Image 80

Chapter 4 Calibration Procedures

AC Amplitude Flatness Adjustment

This procedure can be performed with one of three types of measurement device; a broadband ACrms voltmeter, a power meter, or a thermal voltage converter. The procedure differs slightly depending upon the type of measurement device used. These adjustments us a 50W output termination.

1Use a DMM to measure the ACrms output voltage of the function generator and enter the measurement value for the setup in the table below.

 

Nominal Output

 

 

 

 

 

SETUP

FREQUENCY

AMPLITUDE

Reference for:

64

1 kHz

3.0 V rms

1 kHz flatness DAC gain

 

 

 

 

2a. If you are using a broadband ACrms voltmeter, proceed to step 3.

b.If you are using a power meter capable of measurements at 1 kHz,

use the power meter to measure the function generator output and enter the value for the setup in the table below. (If your power meter does not measure to 1 kHz, see the transfer measurement procedure below.)

 

Nominal Output

 

SETUP

FREQUENCY

AMPLITUDE

Reference for:

 

 

 

 

83

1 kHz

3.0 V rms

V rms, dBm

 

 

 

 

Transfer Measurement Procedure

If you are using a power meter not capable of measurement to 1 kHz, you can perform the transfer measurement at a different frequency. For example, the Agilent E4418A Power Meter with the Agilent 8482A probe and 20 dB attenuator are specified to a low frequency of 100 kHz. To use this measurement device, perform step 1, then use setup 65 to obtain a 100 kHz output. Measure the output with the power meter and record the measured value. Perform setup 83 and enter the recorded value (not a new measurement). Then, perform step 3 (you will use setup 65 twice). This procedure assumes the output of the function generator is flat from 1 kHz to 100 kHz.

c.If you are using a Thermal Voltage Converter (TVC), use the TVC to measure the function generator output and enter the measurement for the setup in the table below (TVC values entered are in mVdc).

 

Nominal Output

 

 

 

 

 

SETUP

FREQUENCY

AMPLITUDE

Reference for:

84

1 kHz

3.0 V rms

Thermal Voltage Converter

 

 

 

 

78

Page 80
Image 80
Agilent Technologies 33120A manual AC Amplitude Flatness Adjustment