Chapter 4 Calibration Procedures

AC Amplitude Flatness Adjustment

3For each setup in the table below, use the CALIBRATE command to change the displayed amplitude to match the measured amplitude.

 

Nominal Output

 

 

 

 

 

 

 

 

 

SETUP

FREQUENCY

AMPLITUDE

Adjustment for:

 

 

65

100 kHz

3.0 V rms

100 kHz amplitude flatness

 

 

 

 

 

 

 

 

66

500 kHz

3.0 V rms

500 kHz amplitude flatness

 

 

 

 

 

 

 

 

67

1 MHz

3.0 V rms

1 MHz amplitude flatness

 

 

 

 

 

 

 

 

68

3 MHz

3.0 V rms

3 MHz amplitude flatness

 

 

 

 

 

 

 

 

69

5 MHz

3.0 V rms

5 MHz amplitude flatness

 

 

 

 

 

 

 

 

70

7 MHz

3.0 V rms

7 MHz amplitude flatness

 

 

 

 

 

 

 

 

71

9 MHz

3.0 V rms

9 MHz amplitude flatness

 

 

 

 

72

10 MHz

3.0 V rms

10 MHz amplitude flatness

 

 

 

 

 

 

 

 

73

10.5 MHz

3.0 V rms

10.5 MHz amplitude flatness

 

4

 

 

 

 

 

74

11 MHz

3.0 V rms

11 MHz amplitude flatness

75

11.5 MHz

3.0 V rms

11.5 MHz amplitude flatness

 

 

 

 

 

 

 

 

 

 

76

12 MHz

3.0 V rms

12 MHz amplitude flatness

 

 

 

 

 

 

 

 

77

12.5 MHz

3.0 V rms

12.5 MHz amplitude flatness

 

 

 

 

 

 

 

 

78

13 MHz

3.0 V rms

13 MHz amplitude flatness

 

 

 

 

 

 

 

 

79

13.5 MHz

3.0 V rms

13.5 MHz amplitude flatness

 

 

 

 

 

 

 

 

80

14 MHz

3.0 V rms

14 MHz amplitude flatness

 

 

81

14.5 MHz

3.0 V rms

14.5 MHz amplitude flatness

 

 

 

 

 

 

 

 

82

15 MHz

3.0 V rms

15 MHz amplitude flatness

 

 

 

 

 

 

 

 

Shift

Menu On/Off

4Perform the Amplitude Flatness Verification procedures on page 60.

Completion of adjustment procedures. Return the function generator to the normal operating mode.