Chapter 4 Calibration Procedures
Performance Verification Tests
Quick Performance Check
The quick performance check is a combination of internal
To perform the quick performance check, do the following:
œSet the function generator to the default power on state (power on default). A procedure is given on page 41.
œPerform a complete
œ Perform only the performance verification tests indicated with | 4 |
the letter Q. |
|
If the function generator fails the quick performance check, adjustment or repair is required.
Performance Verification Tests
The performance verification tests are recommended as acceptance tests when you first receive the function generator. The acceptance test results should be compared against the 1 year test limits. After acceptance, you should repeat the performance verification tests at every calibration interval.
If the function generator fails performance verification, adjustment or repair is required.
55