Chapter 2 Quick Start

To set the output amplitude

To set the output amplitude

At power-on, the function generator outputs a sine wave with an amplitude of 100 mV peak-to-peak (into a 50W termination).

The following steps show you how to change the amplitude to 50 mVrms.

1Enable the amplitude modify mode.

The displayed amplitude is either the power-on value or the previous amplitude selected. When you change functions, the same amplitude is used if the present value is valid for the new function.

100.0 mVPP

2Enter the magnitude of the desired amplitude. 1

Notice that the Num annunciator turns on and “ENTER NUM” flashes on the display, indicating that the number mode is enabled.

50

To cancel the number mode, press Shift Cancel .

3Set the units to the desired value.

The units are selected using the arrow keys on the right side of the front panel. As soon as you select the units, the function generator outputs the waveform with the displayed amplitude. To turn off the flashing digit, move the cursor to the left of the display using the arrow keys.

50.00 mVRMS

1You can also use the knob and arrow keys to enter a number.

26