Chapter 4 Calibration Procedures
Amplitude Flatness Verification
Amplitude Flatness Verification
This test verifies the output amplitude flatness specification at selected frequencies. If you use a TVC (recommended) or a wide band ACrms voltmeter (with a 50W feed through load), perform this procedure as described. If you are using a measurement device that requires a transfer measurement (for example, a power meter), make the transfer in the reference measurement at 100 kHz.
Set the function generator to the first output indicated in the table below and make a reference measurement. Select each function generator output in the table below and adjust the function generator output amplitude until the measured output is at the reference measurement. Compare the amplitude level set on the front panel to the test limits shown in the table. This test is a 50W output termination test.
|
| Agilent 33120A |
| Measurement | ||
| Function | OUT 1 | Ampl | Freq | Nominal | Error |
| TERM | |||||
|
|
|
|
|
| |
Q | Sine wave | 50 W | 3.0 Vrms | 1.00 kHz | <reference> |
|
| Sine wave | 50 W | 3.0 Vrms | 100.00 kHz | <reference> | – 0.03 Vrms |
| Sine wave | 50 W | 3.0 Vrms | 500.00 kHz | <reference> | – 0.045 Vrms |
|
|
|
|
|
|
|
Q | Sine wave | 50 W | 3.0 Vrms | 1.00 MHz | <reference> | – 0.045 Vrms |
|
|
|
|
|
|
|
| Sine wave | 50 W | 3.0 Vrms | 3.00 MHz | <reference> | – 0.06 Vrms |
|
|
|
|
|
|
|
| Sine wave | 50 W | 3.0 Vrms | 5.00 MHz | <reference> | – 0.06 Vrms |
|
|
|
|
|
|
|
| Sine wave | 50 W | 3.0 Vrms | 7.00 MHz | <reference> | – 0.06 Vrms |
| Sine wave | 50 W | 3.0 Vrms | 9.00 MHz | <reference> | – 0.06 Vrms |
| Sine wave | 50 W | 3.0 Vrms | 11.00 MHz | <reference> | – 0.06 Vrms |
|
|
|
|
|
|
|
| Sine wave | 50 W | 3.0 Vrms | 13.00 MHz | <reference> | – 0.06 Vrms |
|
|
|
|
|
|
|
Q | Sine wave | 50 W | 3.0 Vrms | 15.00 MHz | <reference> | – 0.06 Vrms |
|
|
|
|
|
|
|
1Output termination set using front panel controls. HIGH Z assumes no load on output. 50W assumes a 50W
60