Performing Tests
TX Test Port Coupling Measurements
gy A5 Hzyypn—
•Full calibration requires two separate measurement setups. Make certain to use the correct test cables for each setup.
•Partial calibration uses one measurement setup. Make certain to use the Base Station TX Test Port cable when making the measurement.
fyy fy
The Test Software will calculate and display the Base Station TX Test Port coupling factor. You should write this value next to the TX Test Port for future reference.
Press the k1 (Proceed) key to return to the Calibration/Utilities Menu screen, then press the k5 (Return) key to return to the Lucent AMPS Test Main Menu screen.
If you then select the Base Station Configuration field (from the Lucent AMPS Test Main Menu screen), you will note that the TX Test Port Coupling Factor [dB] field for the specified antenna now contains the value determined by the test.
90 | Chapter 4 |