Reference
Parameters
;2 LAP Baud Rate [valid test set range]
=2 LAP Data Length [7 or 8]
>2 LAP Flow Control [1=None 2=Xon 3=Hw]
•If you select 1, the link will use no flow control.
•If you select 2, the link will use Xon/Xoff flow control.
•If you select 3, the link will use hardware flow control. This setting must be used for baud rates of 38400 and higher.
10.LAP Parity [0=0 1=1 2=none 3=even 4=odd]
Enter the selection to determine the parity style that will be used for communication between the PC and the SERIAL 9 port on the Test Set.
•If you select 0, a zero parity will be used.
•If you select 1, a ones parity will be used.
•If you select 2, no parity will be used.
•If you select 3, an even parity will be used.
•If you select 4, an odd parity will be used.
11.LAP Stop Length [1 or 2]
Enter the selection to match the PC communication program settings that will be used for the number of stop bits in communications between the PC and the SERIAL 9 port on the Test Set.
•If you select 1, one stop bit will be used.
•If you select to 2, two stop bits will be used.
12.Not Used
13.Not Used
14.Not Used
15.Not Used
16.Not Used