Step 20. When menu C3 reappears, select START CAL and
press the Enter key to begin the calibration proce-
dure.
Continue the calibration sequence by following the
prompts as they appear. Connect the appropriate Iso-
lation Devices, Broadband Loads, Opens, Shorts, and
Throughlines, when requested in the calibration se-
quence.

37xxxE OM 7-27

MEASUREMENT CALIBRATION SOLT CALIBRATION
MENU C3
CONFIRM
CALIBRATION
PARAMETERS
PORT 1 CONN
TYPE N (M)
PORT 2 CONN
SMA (M)
REFLECTION
PAIRING
MIXED
LOAD TYPE
SLIDING
THROUGH
PARAMETERS
REFERENCE
IMPEDANCE
TEST SIGNALS
START CAL
PRESS <ENTER>
TO SELECT
OR CHANGE