If you are creating a new file, enter the filename and
select DONE when finished.
10-7 USING AUTOCAL An example procedure for using the AutoCal module is given below.
This example assumes a frequency range of 40 MHz to 40 GHz, a
power level of -7 dBm, and use of a Series 36582 AutoCal module.
Step 1. Press the Begin Cal key (top left).
Step 2. Select AUTOCAL from the displayed menu (C11,
left).

37xxxE OM 10-9

AUTOCAL USING AUTOCAL

.-
Measurement
Enhancement
Channels
Display
Begin
Cal
Apply
Cal
MENU C11
BEGIN CALIBRATION
KEEP EXISTING
CAL DATA
REPEAT
PREVIOUS CAL
AUTOCAL
CAL METHOD
XXXXXXXXX
TRANSMISSION
LINE TYPE:
XXXXXXXXX
CHANGE CAL
METHOD AND
LINE TYPE
NEXT CAL STEP
PRESS <ENTER>
TO SELECT