A-106 37xxxE OM

G ALPHABETICAL LISTING
1. TEST PORT 1 POWER SHOULD BE APPROXIMATELY=
AUT(x dB compression spec) - AUT(gain) - 15 dB
2. PORT 2 INPUT POWER SHOULD BE LESS THAN 0dBm
(UNLESS OPTION 6 IS INSTALLED).
3.NOMINAL OFFSET =APPROXIMATE GAIN (ORL
OSS)
OF EXTERNALDEVICES PRECEDINGTHE AUT.
4. DEFAULT DISPLAY IS DUAL CHANNEL1-3 IN WHICH
CHANNEL1 = b2/1 [dBm] AND CHANNEL3 = S21.
1. AFTER THE AUT IS CONNECTED,NORMALIZE S21.
2. INDICATE THE GAIN COMPRESSION POINT VALUE

(

xdB

)

AND SELECT <TEST AUT>.
3. INCREASE TEST PORT 1 POWER UNTIL A 1 dB (or x dB)
DECREASE INS21 IS OBSERVED.
PORT1 PORT2
AMP ATTN TEST PORTS ATTN
12
NOMINAL OFFSET
- SWEPTFREQUENCY GAIN COMPRESSION-
- CALIBRATION INSTRUCTIONS -
- MEASUREMENTINSTRUCTIONS -
Menu EXT_GC1, Gain Compression Help Menu 1