A-30 37xxxE OM

C ALPHABETICAL LISTING
MENU DESCRIPTION GPIB COMMAND
DISCRETE FILL None
INPUT START,
INCR, POINTS,
THEN SELECT
“FILL RANGE”
This menu is used to create one or more ranges of discrete
equally spaced frequency points for calibration. None
START FREQ
XXX.XXXXXXXXX GHz Enter the first frequency of the range. FRS; FRS?
INCREMENT
XXX.XXXXXXXXX GHz Enter the increment (step size) between one frequency and
the next. FRI; FRI
NUMBER OF PTS
XXXX POINT(S) Enter the number of frequency points in the range. FRP; FRP?
STOP FREQ
XXX.XXXXXXXXX GHz Enter the stop frequency, in GHz. None
FILL RANGE
( XXXX ENTERED) Moving the cursor here and pressing ENTER fills the range
and shows the number of frequencies selected (in
NUMBER OF PTS above).
FIL
INDIVIDUAL
FREQ INSERT Calls menu C2A, which allows you to set the individual
frequencies. DFQ
CLEAR ALL Clears all entries displayed above. FRC
FINISHED
NEXT CAL STEP Calls menu C3, the next menu in the calibration sequence. None
PRESS <ENTER>
TO SELECT Pressing the Enter key implements your menu selection. None
Menu C2D, Fill Frequency Ranges