A-224 37xxxE OM

S ALPHABETICAL LISTING
MENU DESCRIPTION GPIB COMMAND
TEST SIGNALS None
EXT SOURCE 1 PWR
+XX.XX dBm Enter and display the power level, in dBm, of the 1st
external frequency source. PWR1; PWR1?
EXT SOURCE 2 PWR
+XX.XX dBm Enter and display the power level, in dBm, of the 2nd,
external frequency source. PW2; PW2?
PORT 1 ATTN
0*10dB(0-70) Attenuates the microwave source power at port 1 from 0 to
70 dB, in 10 dB steps. The power is attenuated before
being applied to Port 1 for a forward transmission or
reflection test (S21 or S11, respectively).
(NO STEP ATTENUATOR IN MODEL 372xxE)
SA1; SA1?
PORT 2 ATTN
0*10dB(0-00) Attenuates from 0 to 40 dB (10 dB steps) the microwave
power being input to Port 2 from the device-under-test
(DUT).
TA2; TA2?
PREVIOUS MENU Returns you to the previous menu. (RESUME CAL may be
used instead of PREVIOUS MENU, when accessed during
a calibration.)
None
PRESS <ENTER>
TO SELECT
OR TURN ON/OFF
Pressing the Enter key implements your menu selection. None
Menu SU2A or CAL_SU2A, Sweep Setup 2A