Step 5. Press the Begin Cal key (below).
Step 6. Follow the menu prompts and choose to perform a
full 12-term calibration. Use the Adapter’s SMA
male connector (Y') as Test Port 1 and the VNA’s Test
Port 2 connector as Test Port Y.
Step 7. Press the Save/Recall Menu key (below).
Step 8. Choose SAVE from the displayed menu (top left).
Step 9. Choose the appropriate SD Card or USB drive loca-
tion, based on individual preference (Menu SR2, bot-
tom left).
Step 10. When prompted, select CREATE NEW FILE and en-
ter a conventional DOS filename, such as YP-
RIME_Y.CAL. (Store this file in the current direc-
tory.)
Step 11. Now connect the Adapter’s SMA male end to the
VNAs Test Port 2 SMA female connector.
ADAPTER REMOVAL MEASUREMENTS

8-36 37xxxE OM

.-