A-16 37xxxE OM

A ALPHABETICAL LISTING
MENU DESCRIPTION GPIB COMMAND
AUTOCAL
FULL 2 PORTS None
ISOLATION
AVERAGING None
OMIT Select to omit the isolation step. ACOMIT
DEFAULT Select to use the Default value during the isolation step. ACDEF
AVERAGING
FACTOR
XXXX
Select for user defined averaging factor during the isolation
step. ACIAF; ACIAF?
THRU TYPE
CALIBRATION/TRUE Switch between the Thru in the AutoCal module
(CALIBRATOR) and your own port-to-port Thru (TRUE) to
be use in the Thru Update (default CALIBRATOR).
ACF2TT;
ACF2TC;
ACF2TX?
PORT CONFIG
L=1, R=2
R=1, L=2
Switch between the side of the AutoCal module which is
connected to Port 1 and Port 2 (default LEFT connected to
Port 1, RIGHT connected to Port 2).
ACL1R2; ACRIL2
TEST SIGNALS Calls menu CAL_SU2, which lets you enter calibrate Flat
Test Port Power or change source power(s) and attenuator
settings.
None
START AUTOCAL Calls menu CAL_SEQ, which starts the AutoCal calibration
sequencing. None
PRESS <ENTER>
TO SELECT
OR SWITCH
Press the Enter key to select or switch. None
Menu ACAL_FULL, AutoCal Full Menu