37xxxE OM A-231
ALPHABETICAL LISTING S
FLAT POWERCALIBRATION ADJUSTS THE SOURCE
OUTPUTPOWERAT EACH MEASUREMENTPOINT ACROSS
A FREQUENCYSPAN TO PROVIDE A CONSTANTPOWER
LEVEL AT THE TEST PORT (FORWARDDIRECTION ONLY).
1. PRESET, ZERO, AND CALIBRATE THE POWERMETER.
2. CREATE AND ACTIVATE THE POWERMETER'S CAL
FACTOR LIST FOR THE POWERSENSORBEING USED.
3. CONNECTTHE POWERMETERTO THE DEDICATED GPIB
INTERFACE AND THE POWERSENSORTO THE TEST PORT.
4. SELECT <START FLAT POWERCALIBRATION>.
- FLAT POWERCALIBRATION -
- INSTRUCTIONS -
000.0 %
Menu EXT_SU8 Flat Power Calibration Instructions