Embedding Embedding of Matching Networks, or other components, can be done
as follows:
Step 1. Press the APPL menu key on the instrument’s front
panel.
Step 2. Select the VNA test port where the network will be
embedded, then select the EMBED function under
the EMBED/DE-EMBED S2P FILE menu (menu
DE8, top left).
Step 3. Select APPLY NETWORK S2P FILE TO CAL
FILE DATA. This will take you to the
EMBED/DE-EMBED S2P FILE menu (menu DE9,
middle left).
Step 4. Select where to read the calibration file from.
The calibration file to be embedded can be stored on
either the USB drive or SD Card. Once the calibra-
tion file is recalled, the EMBED/DE-EMBED S2P
FILE menu (menu DE9A, bottom left) is displayed.
Step 5. Menu DE9A allows you to choose the .S2P file to be
embedded from either the SD Card or USB drive. Se-
lect the .S2P file to read.
Once the .S2P file is selected, it is embedded with
the calibration file. To embed additional files, save
the calibration along with the embeded .S2P file onto
the SD Card or USB drive for later recall upon the
next embedding operation.
37xxxE OM 8-63
MEASUREMENTS EMBEDDING/ DE-EMBEDDING
MENU DE8
EMBED/DE-EMBED
S2P FILE
PORT 1/PORT 2
METHOD
EMBED/DE-EMBED
SWAP PORTS OFF
OF S2P DATA
APPLY NETWORK
S2P FILE DATA TO
CAL FILE DATA
PRESS <ENTER>
TO SELECT
OR CHANGE
MENU DE9
EMBED/DE-EMBED
S2P FILE
ORIGINAL CAL FILE
TO APPLY NETWORK
READ CAL FILE
FROM SD CARD
READ CAL FILE
FROM USB DRIVE
PRESS <ENTER>
TO SELECT
PRESS <CLEAR>
TO ABORT
.-
Measurement
Enhancement
Channels
Display
Appl
MENU DE9A
EMBED/DE-EMBED
S2P FILE
S2P FILE DATA
OF THE NETWORK
READ S2P FILE
FROM SD CARD
READ S2P FILE
FROM USB DRIVE
PRESS <ENTER>
TO SELECT
PRESS <CLEAR>