Flowmeter Startup
Figure 3-2 ProLink II – mA output trim procedure
ProLink Menu
Calibration
Milliamp Trim 1
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
4 mA trim |
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Read mA output at |
|
|
|
|
|
| |||||
|
|
|
|
| receiving device |
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Next |
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Enter receiving device |
|
|
|
|
|
| |||||
|
|
| value in Enter Meas |
|
|
|
| Next | ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Next |
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Read mA output at |
|
|
|
|
|
| |||||
No |
|
|
| receiving device |
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
| Yes |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
| Equal? |
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
20 mA trim
Read mA output at
receiving device
Next
Enter receiving device value in Enter Meas
Next
Read mA output at
No receiving device
Equal?
Yes
Finish
3.5Zeroing the flowmeter
Zeroing the flowmeter establishes the flowmeter’s point of reference when there is no flow. The meter was zeroed at the factory, and should not require a field zero. However, you may wish to perform a field zero to meet local requirements or to confirm the factory zero.
Note: Do not zero the flowmeter if a high severity alarm is active. Correct the problem, then zero the flowmeter. You may zero the flowmeter if a low severity alarm is active. See Section 5.4 for information on viewing transmitter status and alarms.
When you zero the flowmeter, you may need to adjust the zero time parameter. Zero time is the amount of time the transmitter takes to determine its
•A long zero time may produce a more accurate zero reference but is more likely to result in a zero failure. This is due to the increased possibility of noisy flow, which causes incorrect calibration.
•A short zero time is less likely to result in a zero failure but may produce a less accurate zero reference.
The default zero time is 20 seconds. For most applications, the default zero time is appropriate.
You can zero the flowmeter with ProLink II or with the zero button on the transmitter.
If the zero procedure fails, see Section 11.6 for troubleshooting information.
12 | Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application |