Measurement Performance

Figure 10-2D1 and D2 density calibration – ProLink II

Close shutoff valve

downstream from sensor

D1 calibration

Fill sensor with D1 fluid

ProLink Menu >

Calibration >

Density cal – Point 1

Enter density of D1 fluid

Do Cal

Calibration in Progress

light turns red

Calibration in Progress

light turns green

Close

D2 calibration

Fill sensor with D2 fluid

ProLink Menu >

Calibration >

Density cal – Point 2

Enter density of D2 fluid

Do Cal

Calibration in Progress

light turns red

Calibration in Progress

light turns green

Close

Done

Compensation

Measurement Performance

Figure 10-3D3 or D3 and D4 density calibration – ProLink II

Close shutoff valve

downstream from sensor

D3 calibration

 

D4 calibration

Fill sensor with D3 fluid

 

Fill sensor with D4 fluid

ProLink Menu >

 

ProLink Menu >

Calibration >

 

Calibration >

Density cal – Point 3

 

Density cal – Point 4

Enter density of D3 fluid

 

Enter density of D4 fluid

Do Cal

 

Do Cal

Calibration in Progress

 

Calibration in Progress

light turns red

 

light turns red

Calibration in Progress

 

Calibration in Progress

light turns green

 

light turns green

Close

 

Close

 

 

 

Done

Done

 

 

Troubleshooting

Defaults

Configuration and Use Manual

89