Measurement Performance
Figure
Close shutoff valve
downstream from sensor
D1 calibration |
Fill sensor with D1 fluid |
ProLink Menu > |
Calibration > |
Density cal – Point 1 |
Enter density of D1 fluid |
Do Cal |
Calibration in Progress |
light turns red |
Calibration in Progress |
light turns green |
Close |
D2 calibration |
Fill sensor with D2 fluid |
ProLink Menu > |
Calibration > |
Density cal – Point 2 |
Enter density of D2 fluid |
Do Cal |
Calibration in Progress |
light turns red |
Calibration in Progress |
light turns green |
Close |
Done |
Compensation
Measurement Performance
Figure
Close shutoff valve
downstream from sensor
D3 calibration |
| D4 calibration |
Fill sensor with D3 fluid |
| Fill sensor with D4 fluid |
ProLink Menu > |
| ProLink Menu > |
Calibration > |
| Calibration > |
Density cal – Point 3 |
| Density cal – Point 4 |
Enter density of D3 fluid |
| Enter density of D4 fluid |
Do Cal |
| Do Cal |
Calibration in Progress |
| Calibration in Progress |
light turns red |
| light turns red |
Calibration in Progress |
| Calibration in Progress |
light turns green |
| light turns green |
Close |
| Close |
|
| |
| Done | Done |
|
|
Troubleshooting
Defaults
Configuration and Use Manual | 89 |