Emerson Process Management 1500 manual Required Transmitter Configuration, Chapter, 4.1Overview

Models: 1500

1 146
Download 146 pages 48.94 Kb
Page 23
Image 23
Required Transmitter Configuration

Chapter 4

Required Transmitter Configuration

4.1Overview

This chapter describes the configuration procedures that are usually required when a transmitter is installed for the first time. The procedures in this chapter should be performed in the order shown in Figure 4-1.

Figure 4-1Required configuration procedures in order

Before You Begin

Using

Characterize the flowmeter

(Section 4.2)

Configure the channels

(Section 4.3)

Configure measurement units

(Section 4.4)

Configure mA output

(Section 4.5)

Configure discrete outputs(1)

(Section 4.6)

Configure discrete input(1)

(Section 4.7)

Done(2)

(1)Only the input or outputs that have been assigned to a channel need to be configured.

(2)If the meter verification option has been purchased, the final configuration step should be to establish a meter verification baseline (see Section 4.8).

ProLink II

Flowmeter Startup

This chapter provides basic flowcharts for each procedure. For more detailed flowcharts, see the ProLink II flowcharts, provided in Appendix C.

Default values and ranges for the parameters described in this chapter are provided in Appendix A.

For optional transmitter configuration parameters and procedures, see Chapter 6. For configuration of the filling and dosing application, see Chapter 7.

Note: All ProLink II procedures provided in this chapter assume that your computer is already connected to the transmitter and you have established communication. All ProLink II procedures also assume that you are complying with all applicable safety requirements. See Chapter 2 for more information.

Required Configuration

Configuration and Use Manual

15

Page 23
Image 23
Emerson Process Management 1500 manual Required Transmitter Configuration, Chapter, 4.1Overview