Troubleshooting
Table 11-7 Sensor pickoff values
Sensor(1) | Pickoff value |
ELITE Model CMF sensors | 3.4 mV |
|
|
Model D, DL, and DT sensors | 3.4 mV |
|
|
Model F025, F050, F100 sensors | 3.4 mV |
|
|
Model F200 sensors (compact case) | 2.0 mV |
|
|
Model F200 sensors (standard case) | 3.4 mV |
|
|
Model H025, H050, H100 sensors | 3.4 mV |
|
|
Model H200 sensors | 2.0 mV |
|
|
Model R025, R050, or R100 sensors | 3.4 mV |
|
|
Model R200 sensors | 2.0 mV |
|
|
Micro Motion | 0.5 mV |
|
|
CMF400 I.S. sensors | 2.7 mV |
|
|
CMF400 sensors with booster amplifiers | 3.4 mV |
|
|
(1) If your sensor is not listed, contact Micro Motion. See Section 1.8
11.23.3Excessive drive gain
Excessive drive gain can be caused by several problems. See Table
Table
Cause | Possible remedy |
|
|
Excessive slug flow | See Section 11.17. |
|
|
Plugged flow tube | Purge the flow tubes. |
|
|
Cavitation or flashing | Increase inlet or back pressure at the sensor. |
|
|
| If a pump is located upstream from the sensor, increase the distance |
| between the pump and sensor. |
|
|
Drive board or module failure, cracked flow tube, | Contact Micro Motion. See Section 1.8. |
or sensor imbalance |
|
|
|
Mechanical binding at sensor | Ensure sensor is free to vibrate. |
|
|
Open drive or left pickoff sensor coil | Contact Micro Motion. See Section 1.8. |
|
|
Flow rate out of range | Ensure that flow rate is within sensor limits. |
|
|
Incorrect sensor characterization | Verify characterization. See Section 4.2. |
|
|
106 | Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application |