Required Transmitter Configuration
Figure
Newer tag | Older tag |
Figure
Newer tag | Older tag |
Before You Begin
Using ProLink II
Density calibration factors
If your sensor tag does not show a D1 or D2 value:
•For D1, enter the Dens A or D1 value from the calibration certificate. This value is the
•For D2, enter the Dens B or D2 value from the calibration certificate. This value is the
If your sensor tag does not show a K1 or K2 value:
•For K1, enter the first 5 digits of the density calibration factor. In the sample tag in Figure
•For K2, enter the second 5 digits of the density calibration factor. In the sample tag in Figure
If your sensor does not show an FD value, contact Micro Motion customer service.
If your sensor tag does not show a DT or TC value, enter the last 3 digits of the density calibration factor. In the sample tag in Figure
Flowmeter Startup
Required Configuration
Configuration and Use Manual | 17 |